Contact Us   
Search   
Your Position: Home >

Biological fermentation equipment

2017/8/16      view:

Biological fermentation equipment

Control parts for fermentor
No. 
 
Parameter name 
 
Technical indicator 
 

 
PH 
 
1.On-line check and measure, and digitization fully automatic control; 
2.PH, acid and base gang control. Peristaltic pump automatically adjust to feed acid and base; 
3.Control range 2-12pH, measure precision ± 0.01 pH, control precision ± 0.01 pH; 
4. It has auto alarm function. 
5.pH sensor and signal transmission line. . 
 

 
DO control 
 
1.On-line check and measure, DO and rotate speed gang control; 
2.Range 0-100% or 0-200%, precision ± 1%; 
3. It has auto alarm function; 
4. DO sensor and signal transmission line.. 
 

 
Foam control 
 
1.Automatically control and automatically alarm; 
2.Eletrode check, peristaltic pump automatically add defoamer. 
 

 
Feed control 
 
1.Maunally control; 
2.Automatically add according to fixed max. height . 
 

 
Temp. control 
 
1.PID fully digitization signal processing, digitization control settings 
2.Circulating water automatically control temperature 
3.Measure range 0-150 degree, control range 15-45 degree,control precision ± 0.5degree 
4.Pt 100 Ω temp. Sensor 
 

 
Rotation speed 
 
1.Fully automatic control settings, With automatic alarm function 
2.Speed range, please check the drawing, precision ± 1% 
3.It has mechanical agitator, turbine type, vertical impeller 
4.Speed:50-200/400rpm 
 

 
Sterilization 
 
Manually control 
Automatic control(optional)

 
Pressure 
 

Checked by pressure gauge, manually control, range 0-0.4Mpa.

automatic control (optional) 


 
Air flow 
 
Display on the rotormeter. 
automatic adjust(optional)
Remarks: the table is only for reference, if client have difference requirement on the control points, we can also make according to client requirement.